High Frequency alternate polarity radio-wave electrical current to cut
or coagulate tissue during surgery.
- Can be either monopolar or bipolar; both actions require the electrical circuit to be completed
- In monopolar action, the electrical current oscillates between the surgeon’s electrode, through the patient’s body, until it meets the ‘grounding plate’ (positioned underneath the patient’s leg) to complete the circuit.
- In bipolar diathermy, the two electrodes are found on the instrument itself. The current is then passed between the electrodes.
- The main two settings of diathermy are cutting and coagulation.
- Cutting uses a continuous waveform with a low voltage.
- In cutting mode, the electrode reaches a high enough power to vaporise the water content.
- it is able to perform a clean cut but is less efficient at coagulating.
- The cutting mode focuses heat at the surgical site, using sparks being the more focussed way to distribute heat.
- Coagulation uses a pulsed waveform with a high voltage.
- The tip should be held slightly away from the tissue, the sparks are spread over a wider area causing charring rather than cutting.
Monopolar diathermy cautery
Comments
Post a Comment